Table F-2: Level Transducer Block Parameters and Descriptions (continued)
Parameter Index Number Description
VOLUME_LOWER 26 The calculated volume value of the lower product at
the current interface
INTERFACE_DISTANCE 27 The distance to the interface (from the upper reference
point).
UPPER_PRODUCT_THICKNESS 28 The thickness of the upper product (from the surface
value to the interface value).
INTERFACE_LEVEL 29 The current interface level value (from the zero level
reference point and up to the interface).
INTERFACE_LEVELRATE 30 The current velocity at which the interface is moving. A
positive value indicates the interface is moving up.
INTERFACE_SIGNAL_STRENGTH 31 The current signal strength of the interface echo.
PROBE_TYPE 32 Select the type of probe that is mounted to this device.
Use User Defined probe if your probe can not be found
in the list or if you have done modifications to a
standard probe.
PROBE_LENGTH 33 Enter the length of the probe measured from the
device's upper reference point (normally the upper side
of the tank flange) down to the probe end. If a weight is
used it shall not be included in the length.
PROBE_ANGLE 34 Defines the angle compared to the plumb line at which
the device with probe is mounted (0°means that probe
is mounted vertically).
PROBE_END_PULSE_POLARITY 35 This parameter is used for User Defined probes only.
Contact Emerson
™
Service Department for more
information
PROBE_IMPEDANCE 36 This parameter is used for User Defined probes only.
Contact Emerson Service Department for more
information
TCL 37 This parameter is used for User Defined probes only.
Contact Emerson Service Department for more
information
PROPAGATION_FACTOR 38 This parameter is used for User Defined probes only.
Contact Emerson Service Department for more
information
REF_PULSE_AMPL 39 This parameter is used for User Defined probes only.
Contact Emerson Service Department for more
information
GEOM_TANK_HEIGHT 40 Tank Height (R)
GEOM_HOLD_OFF_DIST 41 Hold off distance
VAPOR_DC 42 Enter the dielectric constant for the vapor gas in the
tank. For air at 20 degC and atmospherical pressure the
vapor DC is close to 1, but for high pressure/high
temperature applications the DC may increase and
affect the accuracy of the measurement
Level Transducer Block Reference Manual
April 2019 00809-0100-4530
428 Rosemount 5300 Level Transmitter