Figure 3-25: Clearance Distance between the Probe End and the Chamber Bottom
$
A. A clearance distance of 1 in. (25 mm) between the probe end and the chamber bottom is
recommended.
3.4.1
Mount a centering disc on flexible single/twin lead probe
Prerequisites
Note
When using centering discs made of PTFE, note that the maximum temperature is 482 °F
(250 °C).
Procedure
1. Mount the centering disc at the end of the weight.
Reference Manual Mechanical installation
00809-0100-4530 April 2019
Reference Manual 63