EasyManua.ls Logo

Emerson Rosemount 5300 Series - Page 86

Emerson Rosemount 5300 Series
488 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
L, desired probe length:
/
2. Determine n, the number of middle segments needed for the desired probe length.
See Table 3-16.
n, number of middle segments:
3. Calculate Y, the length of the bottom segment. See Table 3-16.
Y, length of bottom segment:
<
4. Continue as follows:
Length of bottom segment (Y) Action
Y < 0.4 in. (10 mm)
<
Continue with step 7. Do not use the
bottom segment.
Y ≥ 0.4 in. (10 mm)
<
Continue with step 5 and cut the bottom
segment.
Y = 31.5 in. (800 mm)
<
a. Add one extra middle segment to the
calculated n.
b. Continue with step 7.
5. Mark where to cut the bottom segment.
1 2
3
4
5
6
7
8
0
1 2
3
0
<
6. Cut the bottom segment at the mark.
Note
Make sure the bottom segment is fixed while cutting.
Mechanical installation Reference Manual
April 2019 00809-0100-4530
86 Rosemount 5300 Level Transmitter