Figure 3-34: Coaxial Probe with Tube
A. Drain
Solids applications
Pull the probe rope through a suitable anchoring point, e.g. a welded eye and fasten it with
two clamps. It is recommended the probe is slack in order to prevent high tensile loads.
The sag should be at least 1.5 in./10 ft. (10 mm/m) of the probe length.
Figure 3-35: Flexible Single Lead Probe with Two Clamps
A. ≥ 10 mm/m
Mechanical installation Reference Manual
April 2019 00809-0100-4530
92 Rosemount 5300 Level Transmitter