EasyManua.ls Logo

Emerson Rosemount 5300 Series - Page 92

Emerson Rosemount 5300 Series
488 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Figure 3-34: Coaxial Probe with Tube
$
A. Drain
Solids applications
Pull the probe rope through a suitable anchoring point, e.g. a welded eye and fasten it with
two clamps. It is recommended the probe is slack in order to prevent high tensile loads.
The sag should be at least 1.5 in./10 ft. (10 mm/m) of the probe length.
Figure 3-35: Flexible Single Lead Probe with Two Clamps
$
A. ≥ 10 mm/m
Mechanical installation Reference Manual
April 2019 00809-0100-4530
92 Rosemount 5300 Level Transmitter