Rosemount 5300 Series
7-28
REMOVING THE
TRANSMITTER HEAD
1. Loosen the nut that connects the
transmitter housing to the Process Seal.
2. Carefully lift the transmitter head.
3. Make sure the upper surface of the
Process Seal is clean and the
spring-loaded pin at the center of the
Process Seal is properly inserted (the pin
should move back when pushed into the
hole).
4. Attach the protection plug to the Process
Seal.
Nut
Process Seal
Put the protection
plug here!
FLANGE VERSION
NOTE!
Do not remove the Process Seal from the
adapter!
THREADED VERSION
Put the protection
plug here!
Nut
Process Seal
Adapter