EasyManua.ls Logo

Emerson ROSEMOUNT 5300 - Installation of Single Lead Probes in Non-Metallic Vessels

Emerson ROSEMOUNT 5300
324 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Reference Manual
00809-0100-4530, Rev BA
July 2009
3-5
Rosemount 5300 Series
A long stud - 10 in. (250 mm) - is recommended for single flexible probes in a
tall nozzle.
Figure 3-3. A single flexible
probe with a long stud.
NOTE!
For single lead probes, avoid 10-in. (250 mm)/DN250 or larger diameter
nozzles, especially in applications with low dielectric constant. An alternative
is to install a smaller nozzle inside the nozzle.
Installation of Single
Lead Probes in
Non-metallic Vessels
For optimal single lead probe performance in non-metallic (plastic) vessels,
the probe must be mounted with a metal flange, or screwed in to a metal
sheet (d>8 in./200 mm), if the threaded version is used.
Figure 3-4. Mounting in
non-metallic vessels.
Electromagnetic disturbances should be kept to a minimum since they may
affect measurement performance.
Metal flange Ø>2
in./DN50
Metal sheet
Ø>8 in./200 mm
Long Stud
10 in. (250 mm)