165
Reference Manual
00809-0100-4026, Rev HA
Section 7: Service and Troubleshooting
November 2014
Service and Troubleshooting
Table 7-11. Resource Block DETAILED_STATUS with Recommended Action Messages
7.15.2 Transducer block
This section describes error conditions found in the sensor transducer block.
Table 7-12. Transducer Block BLOCK_ERR Messages
Table 7-13. Transducer Block XD_ERR Messages
7.15.3 Analog Input (AI) function block
This section describes error conditions that are supported by the AI Block. Read Ta ble 7 -15 to
determine the appropriate corrective action.
Table 7-14. AI BLOCK_ERR Conditions
.
Condition name Recommended action
LOI transducer block error
1. Restart processor
2. Check display connection
3. Call service center
Sensor transducer block error
1. Restart processor
2. Check Rosemount 5400 cable
3. Call service center
Mfg. block integrity error
1. Restart processor
2. Call service center
Non-volatile memory integrity error
1. Restart processor
2. Call service center
ROM integrity error
1. Restart processor
2. Call service center
Condition name and description
Other
Out of Service: The actual mode is out of service
Condition name and description
Electronics failure: An electrical component failed
I/O failure: An I/O failure occurred
Data integrity error: Data stored in the device is no longer valid due to a non-volatile
memory checksum failure, a data verify after write failure, etc.
Algorithm error: The algorithm used in the transducer block produced an error due to
overflow, data reasonableness failure, etc.
Condition
number
Condition name and description
0Other
1 Block configuration error: the selected channel carries a measurement
that is incompatible with the engineering units selected in XD_SCALE,
the L_TYPE parameter is not configured, or CHANNEL = zero
3 Simulate active: Simulation is enabled and the block is using a
simulated value in its execution