EasyManua.ls Logo

Emerson Rosemount 5400 Series - Nozzle Recommendations and Requirements

Emerson Rosemount 5400 Series
314 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
30
Reference Manual
00809-0100-4026, Rev HA
Section 3: Mechanical Installation
November 2014
Mechanical Installation
3.3.6 Nozzle recommendations and requirements
The Rosemount 5400 Series is mounted on a nozzle by using appropriate flanges. For best
performance, it is recommended that the nozzle meets the following recommendations for
height (L) and diameter:
Figure 3-14. Mounting of the Rosemount 5400 Series Transmitter
Table 3-1. Minimum Nozzle Diameter and Recommended Maximum Nozzle Height for Cone Antennas
Model Antenna / material
L
max
in. (mm)
Min. diameter
in. (mm)
5402
(1)
Cone 2 in. (50 mm) SST 6.1 (155) 2.2 (55)
Cone 3 in. (75 mm) SST 5.5 (140) 2.8 (72)
Cone 4 in. (100 mm) SST 8.5 (215) 3.8 (97)
Cone 2 in. (50 mm) Alloy C-276, Alloy 400 5.5 (140) 2.2 (55)
Cone 3 in. (75 mm) Alloy C-276, Alloy 400 6.5 (165) 2.8 (72)
Cone 4 in. (100 mm) Alloy C-276, Alloy 400 9.6 (240) 3.8 (97)
5401 Cone 3 in. (75 mm) SST Pipe installations only
Cone 4 in. (100 mm) SST 5.5 (140) 3.8 (97)
Cone 6 in. (150 mm) SST 6.9 (175) 5.7 (145)
Cone 8 in. (200 mm) SST 10.2 (260) 7.6 (193)
Cone 3 in. (75 mm) Alloy C-276, Alloy 400 Pipe installations only
Cone 4 in. (100 mm) Alloy C-276, Alloy 400 5.5 (140) 3.8 (97)
Cone 6 in. (150 mm) Alloy C-276, Alloy 400 6.9 (175) 5.7 (145)
Cone 8 in. (200 mm) Alloy C-276, Alloy 400 10.2 (260) 7.6 (193)
L
Minimum diameter
> 0.4 in.
(10 mm)
Minimum diameter
L
L
Minimum diameter
Minimum diameter
L
Process seal antenna Rod antenna Cone antenna Extended cone antenna
> 0.4 in.
(10 mm)

Table of Contents