Advanced Diagnostics Configuration
March 2016
Advanced Diagnostics Configuration
6.12 SMART Meter Verification test results
If the SMART Meter Verification test is manually initiated, the transmitter will make several
measurements to verify the transmitter calibration, sensor calibration, coil circuit health,
and electrode circuits health. The results of these tests can be reviewed and recorded on
the “Calibration Verification Report” on page 137. This report can be used to validate that
the meter is within the required calibration limits to comply with governmental regulatory
agencies.
Depending on the method used to view the results, they will be displayed in either a menu
structure, as a method, or in the report format. When using the HART Field Communicator,
each individual component can be viewed as a menu item. When using the LOI, the
parameters are viewed as a method using the left arrow key to cycle through the results. In
AMS Device Manager, the “Calibration Verification Report” on page 137 is populated with
the necessary data eliminating the need to manually complete the report.
When using AMS Device Manager, there are two possible methods that can be used to print
the report.
Method one involves using the print functionality within the EDDL screen. This print
functionality essentially prints a screen shot of the report. If using a standard DD, then a
screen capture will need to be taken using the “Print Screen” button on the keyboard and
pasting the image into a word document.
Method two involves using the print feature within AMS Device Manager while on the status
screen. This will result in a printout of all of the information stored on the status tabs. Page
two of the report will contain all of the necessary calibration verification result data.
The results are displayed in the order found in the table below. Each parameter displays a
value used in the SMART Meter Verification diagnostic evaluation of the meter health.