D.5 Fischer and Porter Sensors
Connect coil drive and electrode cables as shown in Figure D-5.
D.5.1 Model 10D1418 Sensor to Rosemount 8750W
Transmitter
Figure D-5. Wiring for Fischer and Porter Sensor Model 10D1418 and Rosemount
8750W
A. Electrode connections
B. Coil connections
Refer to Figure D-1 on page 225 for actual terminal block configuration drawing.
Rosemount 8750W Transmitter
Table D-6. Fischer and Porter Model 10D1418 Sensor Wiring Connections
Rosemount 8750W Fischer and Porter Model 10D1418 Sensors
1 L1
2 L2
3 Chassis ground
17 3
18 1
19 2
Do not connect mains or line power to the magnetic flowtube sensor or to the
transmitter coil excitation circuit.
1
2
17
18
19
3
A
3
2
1
U1
U2
G
L1
L2
8
6
7
5
B