Implementing a Universal Transmitter
March 2016
Reference Manual
00809-0300-4750, Rev CA
231
Implementing a Universal Transmitter
D.5.2 Model 10D1419 Sensor to Rosemount 8750W
Transmitter
Connect coil drive and electrode cables as shown in Figure D-6.
Figure D-6. Wiring for Fischer and Porter Sensor Model 10D1419 and Rosemount
8750W
A. Electrode connections
B. Coil connections
Refer to Figure D-1 on page 225 for actual terminal block configuration drawing.
Rosemount 8750W Transmitter
Table D-7. Fischer and Porter Model 10D1419 Sensor Wiring Connections
Rosemount 8750W Fischer and Porter Model 10D1419 Sensors
1 L1
2 L2
3 3
17 3
18 1
19 2
Do not connect mains or line power to the magnetic flowtube sensor or to
the transmitter coil excitation circuit.
1
2
17
18
19
3
3
2
1
A
B
16
17
18
L1
L2