Implementing a Universal Transmitter
March 2016
Reference Manual
00809-0300-4750, Rev CA
239
Implementing a Universal Transmitter
D.6.4 Foxboro Sensor to 8750W Transmitter
Connect coil drive and electrode cables as shown in Figure D-14.
Figure D-14. Generic Wiring for Foxboro Sensors and Rosemount 8750W
Refer to Figure D-1 on page 225 for actual terminal block configuration drawing.
Foxboro
Sensor
Rosemount 8750W
Transmitter
A. Electrodes
B. Coils
C. Any shield
D. Chassis ground
Table D-15. Foxboro Generic Sensor Wiring Connections
Rosemount 8750W Foxboro Sensors
1 L1
2 L2
3 Chassis ground
17 Any shield
18 Black
19 White
Do not connect mains or line power to the magnetic flowtube sensor or to the
transmitter coil excitation circuit.
1
2
17
18
19
3
A
B
White
Black
C
D
L2
L1