Flowing full limit
LOI menu path Diagnostics Advanced Diag Meter Verif Test Criteria
Flowing, Full
Set the test criteria for the flowing, full condition. The factory default for this value is set to
five percent with limits configurable between one and ten percent. This parameter applies
to manually initiated tests only.
Empty pipe limit
LOI menu path Diagnostics Advanced Diag Meter Verif Test Criteria
Empty Pipe
Set the test criteria for the empty pipe condition. The factory default for this value is set to
five percent with limits configurable between one and ten percent. This parameter applies
to manually initiated test only.
Continuous limit
LOI menu path
Diagnostics Advanced Diag Meter Verif Test Criteria
Continual
Set the test criteria for the continuous Smart Meter Verification diagnostic. The factory
default for this value is set to five percent with limits configurable between two and ten
percent. If the tolerance band is set too tightly, under empty pipe conditions or noisy
flowing conditions, a false failure of the transmitter test may occur.
9.11 Run commanded Smart Meter Verification
LOI menu path Diagnostics Advanced Diag Meter Verif Run Meter Ver
The Smart Meter Verification diagnostic will be available if the Smart Meter Verification
Professional (MV) was ordered. If MV was not ordered or licensed, this diagnostic will not
be available. This method will initiate the commanded meter verification test.
9.11.1
Test conditions
LOI menu path Diagnostics Advanced Diag Meter Verif Run Meter Ver
Test Condition
Smart Meter Verification can be initiated under three possible test conditions. This
parameter is set at the time that the sensor baseline or Smart Meter Verification test is
manually initiated.
No flow
Run the Smart Meter Verification test with a full pipe and no flow in the line.
Running the Smart Meter Verification test under this condition provides the
Advanced Diagnostics Configuration Reference Manual
November 2019 00809-0100-8782
100 Rosemount
®
8782 Slurry Magnetic Flow Meter Transmitter with HART Protocol