Table 9-2: Manual Smart Meter Verification Test Parameters
Parameter LOI menu path (Diagnostics
Variables MV Results Manual
Results)
1 Test Condition Test Condition
2 Test Criteria Test Criteria
3 8785 Test Result MV Results
4 Simulated Velocity Sim Velocity
5 Actual Velocity ActualVelocity
6 Velocity Deviation Flow Sim Dev
7 Xmtr Cal Test Result Xmtr Cal Verify
8 Sensor Cal Deviation Sensor Cal Dev
9 Sensor Cal Test Result Sensor Cal
10 Coil Circuit Test Result Coil Circuit
11 Electrode Circuit Test Result Electrode Ckt
Table 9-3: Continuous Smart Meter Verification Test Parameters
Parameter LOI menu path (Diagnostics
Variables MV Results Continual
Res)
1 Continuous Limit Test Criteria
2 Simulated Velocity Sim Velocity
3 Actual Velocity ActualVelocity
4 Velocity Deviation Flow Sim Dev
5 Coil Signature Coil Inductnce
6 Sensor Cal Deviation Sensor Cal Dev
7 Coil Resistance Coil Resist
8 Electrode Resistance Electrode Res
9 mA Expected 4–20 mA Expect
10 mA Actual 4–20 mA Actual
11 mA Deviation AO FB Dev
9.14 Smart Meter Verification measurements
The Smart Meter Verification test will make measurements of the coil resistance, coil
inductance, and electrode resistance and compare these values to the values taken during
the sensor baseline process to determine the sensor calibration deviation, the coil circuit
health, and the electrode circuit health. In addition, the measurements taken by this test
can provide additional information when troubleshooting the meter.