Coil circuit resistance
LOI menu path Manual: Diagnostics → Advanced Diag → Meter Verif →
Measurements → Manual Measure → Coil Resist
Continuous: Diagnostics → Advanced Diag → Meter Verif →
Measurements → Continual Meas → Coil Resist
The coil circuit resistance is a measurement of the coil circuit health. This value is
compared to the coil circuit resistance signature measurement taken during the sensor
baseline process to determine coil circuit health. This value can be continuously monitored
using continuous Smart Meter Verification.
Coil inductance (signature)
LOI menu path
Manual: Diagnostics → Advanced Diag → Meter Verif →
Measurements → Manual Measure → Coil Inductnce
Continuous: Diagnostics → Advanced Diag → Meter Verif →
Measurements → Continual Meas → Coil Inductnce
The coil inductance is a measurement of the magnetic field strength. This value is
compared to the coil inductance signature measurement taken during the sensor baseline
process to determine sensor calibration deviation. This value can be continuously
monitored using continuous Smart Meter Verification.
Electrode circuit resistance
LOI menu path
Manual: Diagnostics → Advanced Diag → Meter Verif →
Measurements → Manual Measure → Electrode Res
Continuous: Diagnostics → Advanced Diag → Meter Verif →
Measurements → Continual Meas → Electrode Res
The electrode circuit resistance is a measurement of the electrode circuit health. This
value is compared to the electrode circuit resistance signature measurement taken during
the sensor baseline process to determine electrode circuit health. This value can be
continuously monitored using continuous Smart Meter Verification.
Actual velocity
LOI menu path
Manual: Diagnostics → Advanced Diag → Meter Verif →
Measurements → Manual Measure → ActualVelocity
Continuous: Diagnostics → Advanced Diag → Meter Verif →
Measurements → Continual Meas → ActualVelocity
The actual velocity is a measurement of the simulated velocity signal. This value is
compared to the simulated velocity to determine transmitter calibration deviation. This
value can be continuously monitored using continuous Smart Meter Verification.