11.4 Replacing a terminal block socket module
Note
This section only applies to sensors with Kx approval codes.
The terminal block socket module is shown in Figure 11-2. To gain access to the socket
module, remove the junction box from the sensor adapter.
Figure 11-2: Socket Module—Terminal Block
A
A. Mounting screws:
• 2X—standard
• 4X—with I.S. divider
11.4.1
Removing a terminal block socket module
1. Disconnect power to the transmitter and the remote cabling connected to the
terminal block.
2. Remove the junction box cover to gain access to the remote cabling.
3. To disconnect the terminal block from the junction box housing, remove the two
mounting screws and the two divider mounting screws (if applicable).
4. Pull up on the terminal block to expose the socket module base.
5. To remove the socket module, loosen the two mounting screws and pull up on the
socket module from the base.
6. When removing the socket module, do not pull on the wires.
11.4.2
Installing a terminal block socket module
1. Insert the new terminal block socket module, press the base into its keyed position,
and tighten the two mounting screws.
Reference Manual
Maintenance
00809-0100-8782 November 2019
Reference Manual 117