2. Connect the terminal block to the junction box housing by tightening the two
mounting screws.
Install the divider with the two mounting screws if applicable.
3. Reconnect remote cabling and power and replace junction box cover.
11.5 Replacing a terminal block with amp clips
Note
This section applies only to sensors with Nx apporval codes.
Figure 11-3: Terminal block with amp clips
A
A. Mounting screws:
2X—standard
4X—with I.S. divider
11.5.1
Removing a terminal block
1. Disconnect power to the transmitter.
2. Remove the junction box cover on the sensor to gain access to the remote cabling
and disconnect the remote cabling connected to the terminal block.
3. To disconnect the terminal block from the junction box housing, remove the two
mounting screws and the two divider mounting screws (if applicable).
4. Pull up on the terminal block to expose the connecting wires.
5. To remove the terminal block, unclip both wire connectors.
Maintenance
Reference Manual
November 2019 00809-0100-8782
118 Rosemount
®
8782 Slurry Magnetic Flow Meter Transmitter with HART Protocol