EasyManua.ls Logo

Emerson Rosemount 8782 - Uninstalled Sensor Tests

Emerson Rosemount 8782
174 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Table 12-11: Test D. Electrode to electrode shield
Test conditions Expected
value
Potential cause Corrective action
Location: installed
Required equipment: LCR
(Set to Resistance and 120
Hz)
Measuring at connections:
18 and 17 = R
1
19 and 17 = R
2
R
1
and R
2
should be
stable
|R
1
–R
2
|
≤300Ω
Unstable R
1
or R
2
values
confirm coated electrode
Shorted electrode
Electrode not in contact with
process
Empty pipe
Low conductivity
Leaky electrode
Process reference ground
not connected properly
Remove coating from sensor
wall
Use bullet-nose electrodes
Repeat measurement
Remove sensor and
complete tests in Uninstalled
sensor tests
Connect process reference
ground per Process reference
connection
Table 12-12: Test E. Electrode to Electrode
Test conditions Expected
value
Potential cause Corrective action
Location: installed
Required equipment: LCR
(Set to Resistance and 120
Hz)
Measuring at connections:
18 and 19
18 and 17 = R
1
19 and 17 = R
2
Should be
stable and
same relative
magnitude of
R
1
and R
2
from
Test D
Unstable R
1
or R
2
values
confirm coated electrode
Shorted electrode
Electrode not in contact with
process
Empty pipe
Low conductivity
Leaky electrode
Process reference ground
not connected properly
Remove coating from sensor
wall
Use bullet-nose electrodes
Repeat measurement
Remove sensor and
complete tests in Uninstalled
sensor tests
Connect process reference
ground per Process reference
connection
To test the sensor, a multimeter capable of measuring conductance in nanoSiemens is
preferred. Conductance is the reciprocal of resistance.
Or:
12.8 Uninstalled sensor tests
Sensor troubleshooting can also be performed on an uninstalled sensor. If test results from
installed sensor tests are inconclusive, the next step is remove the sensor and perform the
tests outlined in this section. Take measurements from the recommended terminal
location, and directly on the electrode head inside the sensor. The measurement
Troubleshooting Reference Manual
November 2019 00809-0100-8782
144 Rosemount
®
8782 Slurry Magnetic Flow Meter Transmitter with HART Protocol

Table of Contents

Other manuals for Emerson Rosemount 8782

Related product manuals