9.2 Meter factors
Meter Factor is a gain adjustment that may be needed to adjust for installation effects
caused by less than ideal straight run piping or if pairing the 8782 with an 8707 instead of
with a MS sensor.
The procedure is very similar to Vortex and Coriolis and is entered as a flow multiplication
factor of the range 0.2 to 1.8. Not all installations will require a Meter Factor and the 8782
will have a stable flow rate without a Meter Factor. If replacing an existing legacy
transmitter, a Meter Factor will enable the 8782 to match the legacy transmitter flow
reading if there is a discrepancy.
Performing an auto zero trim is recommended for best performance on the 8782. If an
auto zero trim is performed at any time after entering a meter factor adjustment, the
meter factor will need to be re-calculated and re-entered post auto zero trim or else
accuracy could be impacted.
Meter Factor = Actual Flow Rate (from 8712H) / Current Flow Rate (From 8782)
Example:
The known actual flow rate is 100 GPM.
A newly installed transmitter reads a flow rate of 1150 GPM after auto zero.
Remedy:
1. Perform a Meter Factor calculation of 100/115 = 0.8696.
2. Enter a Meter Factor of 0.8696 into the transmitter.
Note
Four decimal places are available to avoid rounding errors.
Result:
The transmitter now reads 100 GPM
9.3 Licensing and enabling
All advanced diagnostics are licensed by ordering option code MV or DS1. In the event that
a diagnostic option is not ordered, advanced diagnostics can be licensed in the field
through the use of a license key. Each transmitter has a unique license key specific to the
diagnostic option code. A trial license is also available to enable the advanced diagnostics.
This temporary functionality will be automatically disabled after 30-days or when power to
the transmitter is cycled, whichever occurs first. This trial code can be used a maximum of
three times per transmitter. See the detailed procedures below for entering the license
key and enabling the advanced diagnostics. To obtain a permanent or trial license key,
contact an Emerson representative.
9.3.1
Licensing the diagnostics
1. Power up the transmitter.
2. Verify the software version is 7.1.1 software or later.