Emerson Process Management GmbH & Co. OHG3-18
X-STREAM XE
Short Form Manual
HASXEE-SFM-HS
05/2017
Signal damping (set in DAMPING) allows
smoothing the output signal, but also affects
the response time of outputs and display:
The factory setting is 0 seconds.
The maximum possible t
90
time is limited by
the size of the internal sampling buffer and
the sampling rates of the installed measur-
ing principles/sensors.
Multi-channel units: the value for each
channel must be entered separately.
The last line of the second page (Pressure
Compensation) enables the user
to enter the current ambient pressure
manually, if no pressure sensor is in-
stalled,
or
to view the current pressure, if a sensor
is installed ( INSTALLED OPTIONS).
If no pressure sensor is installed, enter the
current ambient pressure here and adjust
it, when signicant changes take place: this
improves the accuracy of the instrument.
Ch1
Measurement 1of2
Ranges..
Damping..
Linearization..
Cross Interference..
Average..
Delay..
Ch1
Measurement 2of2
Cut-off: None
Pressure Compensation ..
Ch1
Pressure compensation
Manual Pressure: 1013 hPa
Pressure 1013 hPa
Pressure Status Good
3.5.7 Checking the Settings
If "AfterCalCheck" is enabled (On), the cali-
bration procedure adds an additional phase
where it measures and stores the concentra-
tion value shortly after the calibration.
Press left several times to return to SETUP
and open MEASUREMENT.