Alignment of COe sensor assemblyFigure 9-9:
A. Matchmark
B. Straightedge
C. Sensor holder flat
7. With one wrench holding the eductor elbow (N, Figure 9-8) and one wrench on flats
of sensor holder (J), unthread and remove the COe sensor assembly (I). Do not allow
the eductor elbow to turn.
8. Slide the band heater (M) and heater insulator (K) from the sensor holder (J).
Remove eductor
Complete the following steps to remove the eductor from the OCX 8800.
Maintenance and service
224 OCX 8800