4. Unthread and remove solenoid valve base.
Remove EMI filter and terminal block
Complete the following steps to remove the EMI filter and terminal block from the OCX
8800.
1. Disconnect EMI filter wiring (Figure 9-23) at terminal block (3).
Removal/installation of EMI filterFigure 9-23:
A. Brown
B. Blue
C. Green
D. EMI filter
E. Ground wire
F. Terminal block
2. Disconnect EMI filter wiring at AC power input terminal block on electronic stack.
3. Unbolt and remove EMI filter (D) from electronic stack.
4. Remove ground wire (E) from terminal block (F).
5. Unbolt and remove terminal block (F) from electronic stack.
Install LOI module and board
Complete the following steps to install the LOI module and board in the OCX 8800.