Instruction Manual
IM-106-880C, OI
Section 7: Foundation
Fieldbus
April 2017
Pressure Transmitter used to Measure Level in an Open
Tank
12
5
Solution to Situation #1
Table 7-23 lists the appropriate configuration settings, and Figure 7-9 illustrates the
correct function block configuration.
Table 7-23 Analog Input Function Diagram for a Pressure Transmitter used in Level Measurement
(Situation #1)
Parameter
Configured Values
L_TYPE
Indirect
XD_SCALE
0 to 7 psi
OUT_SCALE
0 to 16 ft
Figure 7-10 Function Block Diagram for a Pressure Transmitter used in Level Measurement