Safety Data
April 2017
Appendix A: Safety Data
A.1 Appendix contents
Safety Instructions ........................................................................................................................ page -
233 Safety Data Sheet for Ceramic Fiber Products ................................................................................
page-255
High Pressure Gas Cylinders ................................................................................................ .......... page-
262
ATEX Clarification .......................................................................................................... ................ page-
263
A.2 Safety Instructions
IMPORTANT
SAFETY INSTRUCTIONS FOR THE WIRING AND INSTALLATION OF THIS APPARATUS:
The following safety instructions apply specifically to all EU member states. They should be
strictly adhered to in order to assure compliance with the Low Voltage Directive. Non-EU states
should also comply with the following unless superseded by local or National Standards.
1. Adequate earth connections should be made to all earthing points, internal and external, where
provided.
2. After installation or troubleshooting, all safety covers and safety grounds must be replaced. The
integrity of all earth terminals must be maintained at all times.
3. Mains supply cords should comply with the requirements of IEC227 or IEC245.
4. All wiring shall be suitable for use in an ambient temperature of greater than 75 °C.
5. All cable glands used should be of such internal dimensions as to provide adequate cable anchorage.
6. To ensure safe operation of this equipment, connection to the mains supply should only be made
through a circuit breaker which will disconnect all circuits carrying conductors during a fault situation.
The circuit breaker may also include a mechanically operated isolating switch. If not, then another
means of disconnecting the equipment from the supply must be provided and clearly marked as such.
Circuit breakers or switches must comply with a recognized standard such as IEC947. All wiring must
conform with any local standards.
7. Where equipment or covers are marked with the symbol to the right, hazardous
voltages are likely to be present beneath. These covers should only be removed
when power is removed from the equipment - and then only by trained service
personnel.
8. Where equipment or covers are marked with the symbol to the right, there is a
danger from hot surfaces beneath. These covers should only be removed by trained
service personnel when power is removed from the equipment. Certain surfaces
may remain hot to the touch.
9. Where equipment or covers are marked with the symbol to the right, refer to the
Operator Manual for instructions.
10. All graphical symbols used in this product are from one or more of the following
standards: EN61010-1, IEC417, and ISO3864.
11. Where equipment or labels are marked "Do Not Open While Energized" or similar,
there is a danger of ignition in areas where an explosive atmosphere is present. This equipment should
only be opened when power is removed and adequate time as specified on the label or in the
instruction manual has been allowed for the equipment to cool down - and then only by trained
service personnel.