SPA with HART Alarm
April 2017
Instruction Manual
IM-106-880C, OI
Figure B-5 SPA Front Panel
5. Figure B-6 shows the SPA menu, submenus, and parameter values that must be selected to
configure the SPA for use with the Hazardous Area OCX 8800. Use the following instructions and
selections shown to properly configure the SPA.
a. Press the SELECT pushbutton. Observe the display window on the SPA front
panel. The display window should read SET HART.
b. See the SET HART command in Figure B-6. To the right of the command window
is the SET HART submenu and related parameter values that must be selected via
the front panel pushbuttons on the SPA.
c. In the submenu views shown:
means press the down pushbutton.
means press the up pushbutton.
means press the SELECT pushbutton one time.
means press the or pushbutton until the desired parameter value is
shown in the SPA window.
d. Proceed through the SPA menu, selecting the menu commands and parameter
values indicated. After completing the sequence in the first column, go to the
top of the second column and continue.
e. To exit the menu, repeatedly press SELECT to display any main menu command.
Then, press or until CONF EXIT is displayed. Select CONF EXIT.
f. Detailed instructions concerning the configuration menu and the submenu
structure for each main command are provided in the SPA user's manual.
270 Configuration/Calibration