Instruction Manual
IM-106-880C, OI
Section 3: Configuration and
Startup
April 2017
Section
contents
43
Section 3: Configuration and Startup
3.1 Section contents
Verify Installation ................................................................................................. page
43
Initial Power Up .................................................................................................... page
46
Set Test Gas Values ............................................................................................... page
46
Calibration Solenoids ............................................................................................ page
47
Blowback Feature ................................................................................................. page
48
Calibration Verify Feature ..................................................................................... page
49
Calibration Tolerance Feature ............................................................................... page
52
COe Purge/ Zero Feature ....................................................................................... page
53
OCX 8800 Reset Procedure ................................................................................... page
55
Install all protective equipment covers and safety ground leads after installation. Failure
to install covers and ground leads could result in serious injury or death.
3.2 Verify Installation
Ensure the Hazardous Area OCX 8800 is installed correctly. Verify mechanical installation and all
electrical and pneumatic connections. Refer to Section 1.
Make sure that the Hazardous Area OCX 8800 is turned on and operating prior to firing up
the combustion process. Damage can result from having a cold Hazardous Area OCX
8800 exposed to the process gases.
NOTE:
During outages, and whenever possible, leave all Hazardous Area OCX 8800 units
running to prevent condensation and premature aging from thermal cycling.
3.2.1 Verify Configuration - HART Electronics
There are three switches on the microprocessor board which are user configurable for the