EasyManua.ls Logo

Emerson Rosemount SOLU COMP II - Page 44

Emerson Rosemount SOLU COMP II
58 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
36
MODEL SOLU COMP II SECTION 6.0
CALIBRATION
6.3.2 Procedure — Zeroing the sensor.
1. Place the sensor in the zero standard. Be sure no air bubbles are trapped against the membrane. The sensor
current will drop rapidly at first and then gradually reach a stable zero value. To monitor the sensor current, go
to the main display and press until the sensor input current is showing. Typical zero current for an ozone
sensor is -10 to +10 nA.
A new sensor or a sensor in which the electrolyte solution has been replaced may require several hours (occa-
sionally as long as overnight) to reach a minimum zero current. DO NOT START THE ZERO ROUTINE UNTIL
THE SENSOR HAS BEEN IN ZERO SOLUTION FOR AT LEAST TWO HOURS.
2. Press MENU. The main menu screen appears. Choose Calibrate.
3. Choose Measurement.
4. Choose Zero.
5. The screen at left appears. The top line is the current ozone reading
based on the previous calibration or, for a first time calibration, the default
sensitivity.
6. Once the reading is stable, the screen at left appears. Sensor zero is
complete and the analyzer has stored the zero current. The screen
remains until the operator presses MENU then EXIT to return to the main
display.
NOTE
Pressing ENTER during the zero step will cause the analyz-
er to use the present sensor current as the zero current. If the
sensor is zeroed before the current has reached a minimum
stable value, subsequent readings will be in error.
After zeroing, leave the sensor in the zero solution and verify that the sen-
sor current is between -10 and +10 nA. To display the sensor current, go
the the main display and press until the input current is showing.
7. This screen appears if the zero current is extremely high. See Section 8.3
for troubleshooting. To repeat the zero step, press EXIT and choose
Zero.
8. This screen appears if the zero current is moderately high. To continue,
choose Yes. To repeat the zero step, choose No. See Section 8.3 for
troubleshooting.
CCaalliibbrraattee
Hold
Program Display
CalSensor1?
MMeeaassuurreemmeenntt
Temp
S1 Live 1.000ppm
Zeroing
WWaaiitt
Sensor Zero Fail
Current Too High
S1 Live 0.000ppm
Sensor Zero Done
Cal S1?
InProcess
ZZeerroo
Possible ZeroErr
Proceed? Yes
NNoo

Table of Contents

Related product manuals