TEKNISET TIEDOT
Kuvaus
Valmistaja EMS SA, CH-1260 Nyon, Sveitsi
Malli AIR-FLOW handy 3.0 PERIO
Luokitus 93/42/ETY Luokka IIa
Tila Jatkuva käyttö
Syöttöpaine turbiiniliitäntään:Vesi
1- 2.2 baaria (1000-2200 hPa)
syöttövirtauksella 50-80 ml/min.
Syöttöpaine turbiiniliitäntään:Ilma
Staattinen paine 2.7 - 3.5 baaria
(2700-3500 hPa)
Paino Noin 0,150 kg
Käyttöolosuhteet +10°C - +40°C
suhteellinen kosteus 30 % - 75 %
ilmanpaine 700 hPa - 1060 hPa
Korkeus enintään 3000 m
Käsikappaleen vahvistettu käyttöikä 800 sterilointisykliä
Varastointiolosuhteet -10°C - +40°C
suhteellinen kosteus 10 % - 90 %
ilmanpaine 500 hPa - 1060 hPa
Kuljetusolosuhteet -10°C - +40°C
suhteellinen kosteus 10 % - 90 %
ilmanpaine 500 hPa - 1060 hPa
TECHNICAL DATA
Description
Manufacturer EMS SA, CH-1260 Nyon, Switzerland
Model AIR-FLOW
handy 3.0 PERIO
Classication 93 / 42 EEC Class IIa
Mode Continuous operation
Service Pressure to the turbine
connection : Water
1 to 2.2 bar (1000-2200 hPa)
with a service ow of 50-80 ml/min.
Service Pressure to the turbine
connection : Air
Static pressure 2.7 to 3.5 bar
(2700-3500 hPa)
Weight Approx. 0.150 kg
Operating conditions +10°C to +40°C
30% to 75% relative humidity
700 hPa to 1060 hPa air pressure
Altitude max. 3000 m
Validated life time of the handpiece 800 sterilization cycles
Storage conditions -10°C to +40°C
10% to 90% relative humidity
500 hPa to 1060 hPa air pressure
Transport conditions -10°C to +40°C
10% to 90% relative humidity
500 hPa to 1060 hPa air pressure
Symptom Solutions
Water comes into the powder chamber
or leaks from the cap
Check the dental unit connection
Check the state of the o-rings of the turbine connection
Connect the unit to the dental unit (be careful, the powder chamber must be empty)
Dry and ll the powder chamber
Send the unit to an approved EMS repair center
No powder/air jet coming from the unit
Cut immediately the air input by releasing the foot pedal
Wait for 1-2 minutes for the system to decompress
Disconnect the unit from the dental unit
Press the foot control pedal of the dental unit
if no air comes out of the turbine connector, it is your dental unit that causes the problem
If air comes out, it is the device that causes the problem
Unscrew the cap above the wash-hand basin, because the powder can be ejected. A sealed apparatus, even
disconnected, can indeed remain under pressure
Empty the powder chamber and screw the cap back on
Connect the unit to the dental unit (be careful, the powder chamber must be empty)
Connect the handpiece to the unit
Press the foot control pedal of the dental unit
If air comes through the handpiece connector, the handpiece is blocked. Clean the handpiece
If no air comes out of the handpiece connector, the unit is blocked
Send the unit to an approved EMS repair center
No water spray is coming from the
nozzle
Change the nozzle
Remove the nozzle and verify that a water jet comes out from the handpiece
Send the device to a certied EMS Repair Center
Water leakage between the
handpiece and the nozzle
(unusual leak of droplets)
Check the position of the nozzle and its position on the handpiece
Change the nozzle
Check for residues between the nozzle and the handpiece
Check that nothing a󰀨ects the alignment of the nozzle
Send the device to a certied EMS Repair Center
Water leakage between the nose
connection and handpiece
Check the O-rings on the nose connection of the device, replace them if necessary
Change the handpiece
Send the device to a certied EMS Repair Center
Air and/or powder is leaking from the
threads of the cap
Check the seal and the cleanliness of the thread on the powder chamber and on the cap
Replace the seal or the cap if necessary
Send the device to a certied EMS Repair Center
TROUBLESHOOTING