TEHNIČKI PODATCI
OPIS
Proizvođač EMS ELECTRO MEDICAL SYSTEMS SA,
CH-1260 Nyon, Švicarska
Model AIR-FLOW MASTER PIEZON
®
Klasikacija EN 60601-1 - Klasa I
- Primijenjeni dio, tip B
- IP 20, jedinica
- IP X1, nožna pedala
Klasikacija 93/42/EEZ Klasa IIa
Režim rada
s ispiranjem:
bez ispiranja:
Kontinuirani rad
Ciklus rada 10% najduže tijekom 10 minuta
Napon napajanja 100-240 VAC 50-60Hz
Potrošnja struje maksimalno 100 VA
Osigurači T 2.5 A 250 VAC (Φ 5 X 20 mm, (vremenski,
niske jačine odvajanja)
Dovod vode 1 do 5 bara (100 do 500 kPa)
Dovod komprimiranog zraka 5,5 do 7,5 bara (550 do 750 kPa)
Specikacije ultrazvuka:
Maksimalni izlaz:
Raspon frekvencije
12 watta
24-32 kHz
Težina (približno) 4,5 kg
Dimenzije u mm (VxŠxD) 225 x 280 x 295
Uvjeti rada +10° C do +40° C
30% do 75% relativne vlažnosti
Maksimalna nadmorska visina 3000 m
Uvjeti čuvanja i transporta -10° C do +40° C
10% do 95% relativne vlažnosti
500 hPa do 1060 hPa zračnog tlaka
TECHNICAL DATA
DESCRIPTION
Manufacturer EMS ELECTRO MEDICAL SYSTEMS SA,
CH-1260 Nyon, Switzerland
Model AIR-FLOW MASTER PIEZON
®
Classication EN 60601-1 - Class I
- Applied part, Type B
- IP 20 unit
- IP X1, footpedal
Classication 93/42/EEC Class IIa
Operating mode
with irrigation:
without irrigation:
Continuous operation
Duty cycle 10% during
10 min. maximum
Supply voltage 100-240 VAC 50-60Hz
Power consumption max. 110 VA
Fuses T 2.5 A 250 VAC (Ø5 X 20 mm, (time lag,
low breaking capacity)
Water supply 1 to 5 Bar (100 to 500 kPa)
Compressed air supply 5.5 to 7.5 Bar (550 to 750 kPa)
Ultrasound specications
Maximum output:
Frequency range:
12 Watts
24-32 kHz
Weight (approx.) 4.5 kg
Dimensions in mm (HxWxD) 225 x 280 x 295
Operating conditions +10 °C to +40 °C
30% to 75% relative humidity
Altitude max. 3000 m
Storage and
transport conditions
-10 °C to +40 °C
10% to 95% relative humidity
500 hPa to 1060 hPa air pressure
FB-581_4_Rev A-01_Not Util AIR-FLOW MASTER Piezon.indd 119 08.02.2018 09:56:12