56
A set of interchangeable colored rings is provided to help recognize the chosen powder avors. Please
refer to the information provided in the chapter "Maintenance".
Check the condition of the threads on the powder chamber, on the cap and on the O-ring seals.
Please refer to the information provided in the chapter "Maintenance".
Clean the threads of the powder chamber and of the cap before screwing them on.
Ensure that the powder chambers are closed tight. Otherwise air and powder could leak.
Do not turn the powder chambers upside down after lling.
The PERIO powder being specially ne, tamping it down the powder chamber must be avoided
when lling it.
The powder should not remain over a determined time inside the powder chamber. Read the
instructions delivered with the powders.
The PERIO powder level will go down slightly a few minutes after the lling (powder compaction).