EasyManuals Logo

Endecotts Octagon 200CL Installation Qualification Procedures

Default Icon
2 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Page #1 background imageLoading...
Page #1 background image
ENDECOTTS LIMITED 9 LOMBARD ROAD LONDON SW19 3UP ENGLAND
TEL: +44 (0)20 8542 8121
sales@endecotts.com www.endecotts.com
Reg’d No. 842205 (England) Reg’d Office: Lombard Road, London SW19 3UP
IQOQ-OCTCL200
Page 1 of 2
Installation Qualification and Operational Qualification
Sieve shaker Octagon 200CL
Endecotts sieve shakers are thoroughly tested in our manufacturing facility and are not released for
despatch unless they meet our test criteria.
Installation of the apparatus is described in the Operating Instructions supplied with the apparatus.
This document is intended to aid those users, principally in the pharmaceutical industries, which need
to be able to demonstrate Good Management Practice in the commissioning of new apparatus. The
following checklist should be completed by the user for this purpose.
DO NOT proceed to the next criterion unless you have answered ‘YES’ to the current one.
Numbers in brackets refer to apparatus diagram on page 37 of Operating Instructions.
Operational criteria
Tick
NOTES -
1
I have read and understood the Operating
Instructions supplied with the apparatus.
YES NO
Read the instruction. Contact
Endecotts if you have any query.
2
All listed components (page 24) are present and
packaging has been removed.
YES NO
Check packaging and refer to
Endecotts if not found.
3
Ensure transit bolts are removed (page 27)
YES NO
Operating Instruction - page 27
4
The apparatus is positioned on a suitable level, rigid,
robust bench where it will not affect other
equipment.
YES NO
Operating Instruction - page 27
5
Clamp rods and hand wheels have been assembled.
YES NO
Operating Instruction - page 28
6
Mains electrical supply is as stated on the rating
plate.
YES NO
Operating Instruction - page 26
Place a suitable sieve stack, including lid and receiver, on the location plate and secure
them with the clamping unit
Operating Instruction - page 29
7
The sieve stack and clamp plate are held securely.
YES NO
Operating Instruction - page 29-30
Connect mains lead to electrical supply at rear of apparatus and switch on Mains Switch
Operating Instruction - page 24
8
The "Time" display (3) will illuminate with a value
(in minutes) and the "Amplitude" meter (11) will
illuminate at the factory/previous setting
YES NO
Operating Instruction - page 25
9
Set the sieving time by using "˄" and "˅" (6&8)
underneath the "Time" display - 00:10 - 99:50 min
YES NO
Operating Instruction - page 25
10
Set the Amplitude by using "˄" and "˅" (9&10)
underneath the "Amplitude" display - always start at
the lowest setting
YES NO
Operating Instruction - page 25
11
If required, select the Interval operation by pressing
"M1" interval mode button (2). The indicator LED
will illuminate (1). Switch the Interval function OFF
YES NO
Operating Instruction - page 25
12
Start the machine by pressing the START button "►"
(5). LED light (4) will illuminate
YES NO
Operating Instruction - page 25
13
Vibration commences
YES NO
Operating Instruction - page 25

Other manuals for Endecotts Octagon 200CL

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Endecotts Octagon 200CL and is the answer not in the manual?

Endecotts Octagon 200CL Specifications

General IconGeneral
ModelOctagon 200CL
TypeLaboratory Sieve Shaker
Sieve Diameter200 mm
CapacityUp to 8 sieves
Drive / PrincipleElectromagnetic
Sieve Height50 mm
TimerDigital, 0 - 99 minutes
Intermittent OperationYes
Suitable for Dry SievingYes
Power Supply230 V, 50 Hz
ControlDigital
Suitable for Wet SievingYes
Shaking ActionHorizontal and Vertical

Related product manuals