Creating, editing and saving programs Biometra TRIO
42 Version 03_19
gram preview table. By using the program preview function the template can be
checked before it is opened.
3. Press the [OPEN TEMPLATE] button to open a program template.
Figure 20 Biometra TRIO program template screen
7.3 Assign program names
Each template has a specific name which is superimposed in the programming screen.
To enter a new name touch the input box NAME and enter a new name using the Bi-
ometra TRIO virtual keyboard (see chapter Keyboard"" p. 23).
Note: If a character is pressed, the pre-displayed name will be deleted and a new pro-
gram name can be set. If the program name should be modified please first press an-
other button like delete or space.
7.4 Set heated lid temperature
To set the heated lid temperature, touch the input field LID (see Figure 16) and in the
next screen (see Figure 21) enter a value between 30°C and 110°C using the Biometra
TRIO virtual keypad.
Note: The default setting for the heated lid temperature is 99°C. 99°C is sufficient for
typical PCR programs.
To switch the heated lid on or off use the corresponding buttons.