Biometra TRIO Running, pausing, continuing and stopping programs
Version 03_19 59
Figure 35 Biometra TRIO run screen in spreadsheet view.
The following parameters are summarized in the headline table:
If the heated lid is set to preheating ON (see chapter "Set preheating mode" p. 43),
during the preheating of the heated lid the word “PREHEATING” will be displayed in the
field STEP in the headline table. After the heated lid has reached the target tempera-
ture and the 40s equilibration phase has passed, the first program step starts. During
program run, the current step is highlighted in yellow in GRAPHICAL VIEW and
SPREADSHEET VIEW. In graphical view, each step is divided in heating phase and holding
time and both parts are highlighted separately. Time and temperature increments and
ramping rate adjustments cannot be shown in GRAPHICAL VIEW. Instead the symbols T,
t or ! are displayed (see Figure 34).
9.2 Pause Program
To pause a program press the [PAUSE] button in the Biometra TRIO run screen (see
Figure 34 and Figure 35).