Biometra TRIO Creating, editing and saving programs
47
7.3 Assign program names
Each template has a specific name which is superimposed in the programming screen.
To enter a new name touch the input box Name and enter a new name using the Bi-
ometra TRIO virtual keyboard (see section Keyboard"" p. 26).
Note: If a character is pressed, the pre-displayed name will be deleted and a new pro-
gram name can be set. If the program name should be modified please first press an-
other button like delete or space.
7.4 Set heated lid temperature
To set the heated lid temperature, touch the input field Lid (see Fig. 16) and in the next
screen (see Fig. 21) enter a value between 30°C and 110°C using the Biometra TRIO vir-
tual keypad.
Note: The default setting for the heated lid temperature is 99°C. 99°C is sufficient for
typical PCR programs.
To switch the heated lid on or off use the corresponding buttons.
Fig. 21 Biometra TRIO lid temperature setting screen
NOTICE
The heated lid automatically switches off at a temperature difference of >7
5
°C between
the block and the heated lid
. At these low block temperatures, sample condensation on
the
tube lid is no longer to be expected. In addition, it supports a longer service life of
the Peltier elements in the block, since no unnecessary cooling against the heating cover
is required.