Creating, editing and saving programs Biometra TRIO
52
7.6.5 Set Temperature Optimisation Step (TOS)
For optimisation of new PCR protocols very often gradient enabled Thermocyclers are
used. The Biometra TRIO offers no gradient function but by the temperature optimisa-
tion step (TOS) function programs making use of the three blocks and delivering differ-
ent annealing temperatures at the defined step can be created.
Sample replicates are distributed across the sample blocks and are tested for the tem-
perature giving the optimal result. To program a Temperature Optimisation Step press
the following button.
Temperature Optimisation Step (TOS)
Enter the desired annealing temperature (Annealing Temp.) and define a temperature
increment (Increment) between the sample blocks. The effective temperatures for all
sample blocks are displayed (see Fig. 23).
Note: The annealing temperature is set for sample block 2 and the increment defines
the temperatures for sample block 1 and 3. When setting a positive increment the tem-
perature in sample block 1 is the lowest and the temperature in sample block 3 is the
highest. Use the minus button on the keypad to enter a negative increment. When a
negative value is used as increment the temperature in sample block 1 is the lowest and
the temperature in sample block 3 is the highest.
Fig. 24 Biometra TRIO Temperature Optimisation Step screen
7.7 Insert step
To insert a new step, select a program step in the Biometra TRIO programming screen
(see Fig. 16 and Fig. 17) and press the insert step button.