Tools Biometra TRIO
94
Last Power-down by date and time
Table with failure messages by number, date, time, error code and message
Fig. 68 Biometra TRIO Power-On-Logfile screen
11.3.3 Extended Self-Test-Logfile
During the Extended Self-Test the Biometra TRIO Thermal Cycler checks several func-
tions and components like cooler, thermal tracking, heating and cooling rate, refrigera-
tion, heated lid and regulation (see fig. below). Press on the corresponding button to
see the results for a specific test summarized in tabular format.
Fig. 69 Biometra TRIO Extended Self-Test-Logfile screen
The test result can be “Passed” or “Failed” and is listed in the last row of each table for
each test.