Endress+Hauser Analytik Jena Biometra TRIO - Fig. 72 Biometra TRIO Extended Self-Test Screen; Fig. 73 Biometra TRIO Extended Self-Test Progress Overview Screen

Endress+Hauser Analytik Jena Biometra TRIO
118 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Biometra TRIO Tools
97
Fig. 72 Biometra TRIO Extended Self-Test screen
During the test the software provides a progress overview. If a test is already passed the
result “Failed” or “Passed” is displayed, for the currently performed test “Running” is
shown and all other tests are not finished yet.
Note: The Extended Self-Test will take approximately 30 minutes to finish and the in-
strument is blocked, no other program can be started while the test is running. To stop
the Extended Self-Test press:
Stop Extended Self-Test
If the test is stopped, no corresponding log file will be created.
Fig. 73 Biometra TRIO Extended Self-Test progress overview screen
The results of the Extended Self-Test are summarized and stored by the instrument and
can be viewed by the user (see section "Extended Self-Test-Logfile" p. 94).