EasyManua.ls Logo

Endress+Hauser Cerabar M PMC45 - Page 13

Endress+Hauser Cerabar M PMC45
48 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Cerabar M HART Installation
Endress+Hauser 13
Threaded connection, flush-mounted diaphragm
P01-PMP41xxx-06-09-xx-xx-001
Fig. 9: The flush-mounted version is screwed into the process thread with max. 50 Nm ± 5 Nm (at AF 27).
3.3.2 Installation instructions for devices with diaphragm seals –
PMP46, PMP48
!
Note!
The Cerabar M with a diaphragm seal is screwed in, flanged or clamped, depending on the type
of diaphragm seal.
Together, a diaphragm seal and the pressure transmitter form a closed, calibrated system which
is filled with oil. The filling hole is sealed and should not be opened.
Do not clean or press the diaphragm of the diaphragm seals with hard or pointed objects.
Do not remove diaphragm protection until shortly before installation.
When using a mounting bracket, sufficient strain relief must be ensured for the capillaries in order
to prevent the capillary from buckling (bending radius 100 mm).
Please note that the hydrostatic pressure of the liquid columns in the capillaries can cause zero
point shift. You can correct this zero point shift see Page 29, Section 6.2.5 "Position adjustment
- only display (bias pressure)" (onsite operation) or via communication see Page 33,
Section 6.3.7 "Position adjustment - only display (bias pressure)".
Please note the application limits of the diaphragm seal filling oil as detailed in the Technical
Information for Cerabar M TI399P, "Planning instructions for diaphragm seal systems" section.
In order to obtain more precise measurement results and to avoid a defect in the device, mount the
capillaries as follows:
Vibration-free (in order to avoid additional pressure fluctuations)
Not in the vicinity of heating or cooling pipes
Insulate if the ambient temperature is below or above the reference temperature
With a bending radius of 100 mm.
21
G ½
ø26
G ½
ø50
13
21
max. Druckfestigkeit
100 bar
–Einschweadapter
Bestell-Nr.:52002643
mit3.1 Abnahmeprüfzeugnis
Bestell-Nr.:52010172
Viton-Dichtung
O-Ring 14 x 1,78
Viton oder NBR
SW 27

Related product manuals