Cerabar PMC11, PMC21, PMP11, PMP21, PMP23 Electrical connection
Endress+Hauser 21
0 to 10 V output
Device M12 plug Valve plug Cable
PMC11
PMP11
L–
L+
2
1
3
0.5A
A0017576
L–
L+
1
2
3
4
0.5A
A0022822
-
6.1.2 Supply voltage
L
WARNING
Supply voltage might be connected!
Risk of explosion!
When using the measuring device in hazardous areas, installation must comply with
the corresponding national standards and regulations as well as the Safety Instructions.
All explosion protection data are given in separate documentation which is available
upon request. The Ex documentation is supplied as standard with all devices approved
for use in explosion hazardous areas.
Electronic version Device Supply voltage
4 to 20 mA output PMC11
PMP11
PMC21
PMP21
PMP23
10 to 30 V DC
0 to 10 V output PMC11
PMP11
12 to 30 V DC
6.1.3 Current consumption and alarm signal
Number of wires Device Normal operation Alarm signal
1)
2 PMC11
PMP11
PMC21
PMP21
PMP23
≤ 26 mA > 21 mA
3 PMC11
PMP11
< 12 mA 11 V
1) For MAX alarm (factory setting)
6.2 Connection conditions
6.2.1 Cable specification
For valve plug: < 1.5 mm
2
(16 AWG) and Ø3.5 to 6.5 mm (0.14 to 0.26 in)
6.3 Connection data