EasyManua.ls Logo

Endress+Hauser Cerabar S PMP71 with MID Part Certificate - HistoROM;M-DAT (Optional)

Endress+Hauser Cerabar S PMP71 with MID Part Certificate
60 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Cerabar S PMP71 with 4 to 20 mA HART Operation
Endress+Hauser 27
5.4.4 Taking pressure applied at device as value
Example: configuring upper-range value – assigning 20 mA to the pressure value 400 mbar (6 psi).
5.5 HistoROM
®
/M-DAT (optional)
HistoROM
®
/M-DAT is a memory module which is attached to the electronic insert and fulfills the
following functions:
Back-up copy of configuration data
Copying configuration data from one transmitter to another transmitter
Cyclic recording of pressure and sensor-temperature measured values
Recording diverse events, such as alarms, configuration changes, counters for measuring range
undershooting and overshooting for pressure and temperature, overshooting and undershooting
the user limits for pressure and temperature, etc.
#
Warning!
Detach HistoROM
®
/M-DAT from the electronic insert or attach it to the insert in a de-energized
state only.
P01-xxxxxxxx-19-xx-xx-xx-031
Use "E" to save the new value and exit the editing mode.
See next graphic.
P01-xxxxxxxx-19-xx-xx-xx-032
The new value for the damping is now 30.0 s.
Jump to the next parameter with "E" .
You can get back to the editing mode with "+" or "–".
Onsite display Operation
P01-xxxxxxxx-19-xx-xx-xx-035
The bottom line on the onsite display shows the pressure
present, here 400 mbar (6 psi).
P01-xxxxxxxx-19-xx-xx-xx-036
Use "+" or "–" to switch to the "Confirm" option. The active
option is highlighted in black.
P01-xxxxxxxx-19-xx-xx-xx-037
Use "E" to assign the value (400 mbar (6 psi)) to the
GET URV parameter. The device confirms the calibration
and jumps back to the parameter, here GET URV (see next
graphic).
P01-xxxxxxxx-19-xx-xx-xx-035
Switch to the next parameter with "E" .
Onsite display Operation