Troubleshooting Cerabar S PMP71 with 4 to 20 mA HART
42 Endress+Hauser
710 (W710) Warning
C
B>Set span too small. Not
allowed.
– Values for calibration (e.g. lower-
range value and upper-range value)
are too close together.
– Adjust calibration to suit sensor. (See
also Operating Instructions
BA00413P, parameter description
MINIMUM SPAN or these
Operating Instructions)
51
– The sensor was replaced and the
customer-specific configuration does
not suit the sensor.
– Adjust calibration to suit sensor.
– Replace sensor with a suitable
sensor.
– Unsuitable download carried out. – Check configuration and perform
download again.
711 (A711) Alarm
B
B>LRV or URV out of edit limits – Lower-range value and/or upper-
range value undershoot or overshoot
the sensor range limits.
– Reconfigure lower-range value and/
or upper-range value to suit the
sensor. Pay attention to position
factor.
37
– The sensor was replaced and the
customer-specific configuration does
not suit the sensor.
– Reconfigure lower-range value and/
or upper-range value to suit the
sensor. Pay attention to position
factor.
– Replace sensor with a suitable
sensor.
– Unsuitable download carried out. – Check configuration and perform
download again.
713 (A713) Alarm
B
B>100% POINT level out of edit
limits
– The sensor was replaced. – Carry out calibration again. 39
715 (E715) Error
C
Factory
setting:
Warning
C>Sensor over temperature – The temperature measured in the
sensor is greater than the upper
nominal temperature of the sensor.
(See also Operating Instructions
BA00413P, parameter description
Tmax SENSOR or these Operating
Instructions)
– Reduce process temperature/
ambient temperature.
32
– Unsuitable download carried out. – Check configuration and perform
download again.
716 (E716) Error
B
Factory
setting:
Alarm
B>Process isolating diaphragm
broken
–Sensor defective. –Replace sensor.
– Reduce pressure.
24
717 (E717) Error
C
Factory
setting:
Warning
C>Transmitter over temperature – The temperature measured in the
electronics is greater than the upper
nominal temperature of the
electronics (+88 °C (+190 °F)).
– Reduce ambient temperature. 34
– Unsuitable download carried out. – Check configuration and perform
download again.
718 (E718) Error
C
Factory
setting:
Warning
C>Transmitter under temperature – The temperature measured in the
electronics is less than the lower
nominal temperature of the
electronics
(–43 °C (-45 °F)).
– Increase ambient temperature.
Insulate device if necessary.
35
– Unsuitable download carried out. – Check configuration and perform
download again.
719 (A719) Alarm
B
B>Y-VAL of lin. table out of edit
limits
– At least one Y-VALUE in the
linearization table is below the MIN.
TANK CONTENT or above the
MAX. TANK CONTENT.
– Carry out calibration again. (See also
Operating Instructions BA00413P,
Section 5 or these Operating
Instructions)
40
Code Error type/
NA 64
Message/description Cause Measure Priority