6.6 Locking/unlocking
After all parameters have been entered, the manufacturer's-specific parameters can be
locked.
via the keys +Z, –Z, +S and –S or
via the parameter "Security Looking" by entering a three digit code 130.
(130 is the code for unlocking the parameters)
This protects the measuring point from accidental and unauthorised entries.
Keys
# Key Entry
1 Lock operation:
Press +Z and -S simultaneously once
2 Unlock operation:
Press -Z and +S simultaneously once
Parameter
# Parameter Entry Significance
1 Security
Looking
e.g. 131 Parameters
locked
2 Security
Looking
130 Parameters
unlocked
The table below summarises the locking function.
+
Z
S
+
+
+
Z
S
+
Z
S
+
Z
S
+
+
BA211Y42
Locking
Unlocking
Locking with keys has priority
Locking via Display/
reading of
parameters
Changing/writing via Unlocking via
keys communication keys communication
Keys
yes no no yes no
Parameter
yes no no yes yes
6 Level Measurement Cerabar S FF
44 Endress+Hauser