Installation Ceraphant PTP31B, PTP33B
18 Endress+Hauser
5.5 Post-installation check
Is the device undamaged (visual inspection)?
Does the device conform to the measuring point specifications?
For example:
Process temperature
Process pressure
Ambient temperature range
Measuring range
Are the measuring point identification and labeling correct (visual inspection)?
Is the device adequately protected from precipitation and direct sunlight?
Are the securing screws tightened securely?
Does the pressure compensation element point downwards at an angle or to the side?
To prevent the penetration of moisture: are the connecting cables/plugs pointing downwards?