Description of device parameters Deltabar FMD71, FMD72
56
Reset peakhold (161)
Navigation Expert → Diagnosis → Reset peakhold (161)
Write permission Service engineers/Expert
Description You can reset the peak indicators with this parameter.
Options Cancel
Pressure HP
Pressure LP
Sensor Temp. HP
Sensor Temp. LP
All
Factory setting Cancel
Operating hours (162)
Navigation Expert → Diagnosis → Operating Hours (162)
Write permission No write permissions. Parameter is read only.
Description Displays the hours of operation. This parameter cannot be reset.
Damping switch (164)
Navigation Expert → Measurement → Basic setup → Damping switch (164)
Write permission No write permissions. Parameter is read only.
Description Displays the switch position of DIP switch 2 which is used to switch the damping of the
output signal on and off.
Display Off
The output signal is not damped.
On
The output signal is damped. The attenuation constant is specified in the "Damping"
(017) (184) parameter
Factory setting On
Alarm cur.switch (165)