Deltabar FMD71, FMD72 Description of device parameters
45
Navigation Expert → System → Instrument info → Sens. Ser. No HP (122)
Write permission No write permissions. Parameter is read only.
Description Displays the serial number of the sensor module HP (11 alphanumeric characters).
Enter reset code (124)
Navigation Expert → System → Management → Reset (124)
Write permission Operators/Service engineers/Expert
Description Reset parameters completely or partially to the factory values or order configuration by
entering a reset code, see "Resetting to factory settings (reset)" section.(→  13)
Factory setting 0
Press. Eng. Unit (125)
Navigation Expert → Measurement → Basic setup → Press. Eng. Unit (125)
Write permission Operators/Service engineers/Expert
Description Select the pressure engineering unit. If a new pressure engineering unit is selected, all
pressure-specific parameters are converted and displayed with the new unit.
Options mbar, bar
mmH2O, mH2O,
in, H2O, ftH2O
Pa, kPa, MPa
psi
mmHg, inHg
kgf/cm
2
Factory setting mbar, bar or psi depending on the sensor module nominal measuring range, or as per
order specifications
Temp. eng. unit (126)
Navigation Expert → Measurement → Basic setup → Temp. eng. unit (126)
Write permission Service engineers/Expert
Description Select the unit for the temperature measured values.