Deltabar FMD71, FMD72 Operation options
Endress+Hauser 39
Reset code
1)
Description and effect
62 PowerUp reset (warm start)
The device is restarted.
Data is read back anew from the EEPROM (process is reinitialized).
Any simulation which may be running is ended.
333 User reset
This code resets all the parameters apart from:
- Device tag (022)
- Linearization table
- Operating hours (162)
- Event logbook
- Curr. trim 4 mA (135)
- Curr. trim 20 mA (136)
- Lo trim sensor (131)
- Hi trim sensor (132)
- Lo trim sensor (277)
- Hi trim sensor (278)
Any simulation which may be running is ended.
The device is restarted.
7864 Total reset
This code resets all the parameters apart from:
- Operating hours (162)
- Event logbook
- Lo trim sensor (131)
- Hi trim sensor (132)
- Lo trim sensor (277)
- Hi trim sensor (278)
Any simulation which may be running is ended.
The device is restarted.
1) To be entered in "System" → "Management" → "Reset" (124)
After a "Total reset" in FieldCare you have to press the "refresh" button in order to
ensure that the measuring units are also reset.