EasyManua.ls Logo

Endress+Hauser Deltabar FMD72 - Page 54

Endress+Hauser Deltabar FMD72
134 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Commissioning Deltabar FMD71, FMD72
54 Endress+Hauser
Description
5 Select a level unit via the "Height unit (026)"
parameter, here "m" for example.
Menu path: Setup → Extended setup → Level →
Height unit
h
p
v
h
i
v
C
D
E
F
G
A0034737
C See table, step 10.
D See table, step 8.
E See table, step 9.
F See table, step 11.
G See table, step 12.
h Height
i Current value
p Pressure
v Volume
6 Select the "Wet" option via the "Calibration
mode (027)" parameter.
Menu path: Setup → Extended setup → Level →
Calibration mode
7 Vessel is filled up to 0.5 m (1.6 ft), (49 mbar
(0.72 psi)).
Enter the volume value for the lower
calibration point via the "Empty calib. (028)"
parameter, here 0 liters for example.
Menu path: Setup → Extended setup → Level →
Empty calib.
8 Vessel is filled up to 4.5 m (15 ft), (441 mbar
(6.40 psi)).
Enter the volume value for the upper
calibration point via the "Full calib. (031)"
parameter, here 1 000 l (264 gal) for example.
Menu path: Setup → Extended setup → Level →
Full calib.
9 Enter the density of the calibration medium in
"Adjust density (034)", here 1 g/cm3 (1 SGU),
for example.
Menu path: Setup → Extended setup → Level →
Adjust density
10 Set the volume value for the lower current
value (4 mA) via the "Set LRV (166)" parameter
(0 l).
Menu path: Setup → Extended setup → Current
output → Set LRV
11 Set the volume value for the upper current
value (20 mA) via the "Set URV (167)"
parameter 1 000 l (264 gal).
Menu path: Setup → Extended setup → Current
output → Set URV
12 If the process uses a medium other than that
on which the calibration was based, the new
density must be specified in the "Process
density (035)" parameter.
Menu path: Setup → Extended setup → Level →
Process density
13 Result:
The measuring range is configured for
0 to 1 000 l (0 to 264 gal).
For this level mode, the measured variables %, level, volume and mass are available,
see "Unit before lin. (025)" →  86.
9.10.6 "In height" level selection Calibration without reference
pressure (dry calibration)
Example:
In this example, the volume in a tank should be measured in liters. The maximum volume
of 1 000 l (264 gal) corresponds to a level of 4.5 m (15 ft). The minimum volume of 0

Table of Contents

Other manuals for Endress+Hauser Deltabar FMD72

Related product manuals