Deltabar FMD71, FMD72 Description of Device Parameters
Endress+Hauser 85
LP
Sensor
calibration LP
Sensor trim LP
Pressure
simulation LP
Corrected
pressure
Pressure after
damping
Measured
differential
pressure
Differential
pressure +
pressure
inversion
Position
adjustment
(calib. offset)
Differential
pressure
simulation
Damping P
HP
Sensor
calibration HP
Sensor trim
HP
Pressure
simulation HP
Sensor
pressure HP
Measured
pressure HP
14.4 Setup → Extended Setup
Code definition (023)
Navigation Setup → Extended setup → Code definition (023)
Write permission Operators/Service engineers/Expert
Description Use this function to enter a release code with which the device can be unlocked.
Options A number from 0 to 9999
Factory setting 0
Device tag (022)
Navigation Setup → Extended setup → Device tag (022)
Write permission Operators/Service engineers/Expert
Description Enter the device tag e.g. TAG number (max. 32 alphanumeric characters).
Factory setting No entry or according to order specifications
Operator code (021)
Navigation Setup → Extended setup → Operator code (021)