Description of Device Parameters Deltabar FMD71, FMD72
86 Endress+Hauser
Write permission Operators/Service engineers/Expert
Description Use this function to enter a code to lock or unlock operation.
User entry
To lock: Enter a number ¹ the release code (value range: 1 to 9999).
To unlock: Enter the release code.
Note The release code is "0" in the order configuration. Another release code can be defined in
the "Code definition" parameter. If the user has forgotten the release code, the release code
can be visible by entering the number "5864".
Factory setting 0
14.5 Setup → Extended Setup → Level ("Level" measuring
mode)
Level selection (024)
Navigation Setup → Extended setup → Level → Level selection (024)
Write permission Operators/Service engineers/Expert
Description Select the method for calculating the level
Options In pressure
If this option is selected, specify two pressure/level value pairs. The level value is
displayed directly in the unit that you select via the "Unit before lin." parameter.
In height
If this option is selected, specify two height/level value pairs. From the measured
pressure, the device first calculates the height using the density. This information is then
used to calculate the level in the "Unit before lin." selected using the two value pairs
specified.
Factory setting In pressure
Unit before lin. (025)
Navigation Setup → Extended setup → Level → Unit before lin. (025)
Write permission Operators/Service engineers/Expert
Description Select the unit for the measured value display for the level before linearization.
Example Current measured value: 0.3 ft
New output unit: m
New measured value: 0.3 m