Description of Device Parameters Deltabar FMD71, FMD72
94 Endress+Hauser
Description Set current output if sensor module limits are exceeded or not reached.
Options Warning
The device continues to measure. An error message is displayed.
Alarm
The output signal assumes a value that can be defined by the "Output fail mode (190)/
(051)" function.
Special
The lower sensor module limit is undershot (sensor module LP or HP or complete
system):
Current output = 3.6 mA
The upper sensor module limit is overshot (sensor module LP or HP or complete
system):
Current output assumes a value of 21 - 23 mA, depending on the setting of the "High
alarm curr." (052) parameter.
Factory setting Warning
Alarm cur. switch (165)
Navigation Setup → Extended setup → Current output → Alarm cur.switch (165)
Write permission No write permissions. Parameter is read only.
Description Displays the switching state of DIP switch 3 "SW/Alarm min."
Display SW setting
The alarm current has the value defined in "Output fail mode" (051).
Alarm min.
The alarm current is 3.6 mA, regardless of the software setting.
Output fail mode (051)/(190)
Navigation Setup → Extended setup → Current output → Output fail mode (051)/(190)
Write permission Operators/Service engineers/Expert
Description Select Output fail mode. In case of an alarm, the current and the bargraph assume the
current value specified with this parameter.
Options Max: can be set from 21 to 23 mA
Hold: last measured value is held.
Min: 3.6 mA
Factory setting Max (22 mA)
High. alarm curr. (052)